calibration gas set
Calibration gas set for the ODOR handy
Calibration gas set
All devices with electro-chemical sensors must be calibrated regularly. DVGW refers expressly to this once more in the G280.
Such calibration work, however, should always be performed more or less at the same time as the measurement, so that this is performed by users themselves, whenever possible.
This is done quickly and simply for the ODOR handy by means of the calibration gas set:
1. Connect gas
2. Wait until the value is stable
3. Reset the display value to the actual content using a single adjusting potentiometer
4. Finished
Calibration gas sets are available with THT, mercaptan or mecaptan blend fillings.
The set consists of the following components:
- Light metal pressurised gas bottle
- Filling
- Certification to ISO 6326T2
- Special low-adsorption pressure reducer
- Connection hose
- Instructions
After being filled, each bottle is gauged singly with a gas chromatograph. The measurement conforms to the method specified in ISO 6326T2 (DIN 51855/7).
Odorant content: Accuracy and stability are recorded on each certificate. Should you have further questions, please call us or send an e-Mail to your contact person Frank Sasse (Dipl.-Ing.).